5. Advanced operation
Display function Shows the current display
function.
Ambient Light Compensation Shows the ambient light
compensation status.
Current DICOM S ettings
Reading room
Shows the selected reading room.
To retrieve the DICOM status report:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Select DICOM status report to make the information visible on the screen.
5.21.3 DICOM compliance check
About D ICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Select DICOM compliance check to start the compliance check.
Warning:Pressing a key during the compliance check will abort the check.
5.21.4 DICOM calibration
About DICOM c alibration
TheDICOMcalibrationwilladdacorrectiontothecurrentDICOMcurvetoapproachtheperfectDICOM
curve as good as possible.
To start DICOM calibration:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Select DICOM calibration to start the calibration.
Warning:Pressing a key during calibration will abort the calibration, previous values will be restored.
Note: After calibration, the compliance check will start automatically.
5.21.5 Reset DICOM calibration
About reset DICOM calibration
It is possible to restore the original (not corrected) DICOM curve.
To reset the DICOM calibration:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Enter the DICOM preferences submenu.
4. Select reset DICOM calibration to restore the original (not corrected) DICOM curve.
K5902055 CORONIS FUSION 6MP 14/11/2014
29